Plug the pH probe into the right port before starting the calibration process
To enter the pH calibration menu, press continuously for a few seconds on the key.
In order to cancel the pH calibration process, press the key anytime during the calibration process.
Proceed with instructions, by placing the pH probe in the pH 7 solution, and then press on the key.
Wait for the value to stabilize for at least 10-15 seconds. Stir the probe inside the solution to speed up the process
Press the key when the value doesn't change anymore
Remove the probe from the pH 7 solution, rince with water, and place it in the pH 4 solution. Press the key to continue.
Wait for the value to stabilize for at least 10-15 seconds. Stir the probe inside the solution to speed up the process
Press the key when the value doesn't change anymore in order to finish the pH calibration process